Abstract
Purpose: To develop and implement an acceptance procedure for the new Elekta Unity 1.5 T MRI-linac. Methods: Tests were adopted and, where necessary adapted, from AAPM TG106 and TG142, IEC 60976 and NCS 9 and NCS 22 guidelines. Adaptations were necessary because of the atypical maximum field size (57.4 × 22 cm), FFF beam,