Comparison of thromboelastometry by ROTEM ® Delta and ROTEM ® Sigma in women with postpartum haemorrhage.
Gillissen, Ada; van den Akker, Thomas; Caram-Deelder, Camila; Henriquez, Dacia D.C.A.; Bloemenkamp, Kitty W.M.; Eikenboom, Jeroen; van der Bom, Johanna G.; de Maat, Moniek P.M.
(2019) Scandinavian Journal of Clinical and Laboratory Investigation, volume 79, issue 1-2, pp. 32 - 38
(Article)
Abstract
Haemostatic treatment in women experiencing postpartum haemorrhage is increasingly based on point-of-care devices such as ROTEM ® thromboelastometry. Recently, a fully automated successor of the ROTEM ® Delta device, the ROTEM ® Sigma was introduced. To determine whether these devices provide similar results, we compared ROTEM ® parameters using the
... read more
ROTEM ® Delta and Sigma devices in women experiencing postpartum haemorrhage. Prospective observational cohort study of 23 women experiencing postpartum haemorrhage. ROTEM ® INTEM, EXTEM, FIBTEM and APTEM measurements handled by the ROTEM ® Delta and Sigma devices were compared. ROTEM ® FIBTEM values were also related to Clauss fibrinogen values. A correlation of Spearman’s r (r s ) varying between 0.76 and 0.95 was displayed between clot firmness measured in millimeters at 5 (A5), 10 (A10) and 20 (A20) minutes after start of clot formation measured by EXTEM, INTEM and APTEM assays executed on both devices; A5, A10 and A20 of FIBTEM correlated less well (r S between 0.71 and 0.74), especially after five and ten minutes. Correlation between both devices regarding clotting time (CT) was poor. The observed correlation between levels of Clauss fibrinogen and FIBTEM A5 was r s = 0.70, (95% confidence interval (CI): 0.38 to 0.87) for Delta and r s = 0.85, (CI 0.65 to 0.94) for Sigma. A5, A10 and A20 measured in EXTEM, INTEM and APTEM obtained from ROTEM ® Delta and Sigma devices were similar. EXTEM, FIBTEM and APTEM CT values from both devices showed no correlation. Substantial variation was found between FIBTEM assays of the devices. Consequently, results of FIBTEM assays should always be interpreted in the context of device-specific reference values. Correlation with Clauss fibrinogen was better in the ROTEM ® Sigma device.
show less
Download/Full Text
Keywords: Adult, Blood Coagulation, Female, Fibrinogen/metabolism, Humans, Point-of-Care Systems, Postpartum Hemorrhage/blood, Pregnancy, Prospective Studies, Thrombelastography/instrumentation, Comparative Study, Journal Article, Observational Study
ISSN: 0036-5513
Publisher: Informa Healthcare
Note: Funding Information: We would like to thank scientific laboratory technician D. Priem- Visser, research nurses C. Kolster-Bijdevaate, M.S. Bourgonje-Verhart, C.E. Bleeker-Taborh, E. Roos-van Milligen, R.J. M. Berkhout, E. Sucu, E. C. Willems of Brilman-Tuinhof de Moed and M. Stigter-Dekker and medical students M. van de Sande, R.H. Wouters, and L.S. Smits and the clinical midwifes of the Leiden University Medical Center for their contributions to the TeMpOH-2 study. Publisher Copyright: © 2019, © 2019 The Author(s). Published by Informa UK Limited, trading as Taylor & Francis Group.