Biomechanical effects of a titanium intervertebral cage as a stand-alone device, and in combination with locking plates in the canine caudal cervical spine
Beishuizen, Rick; Reints Bok, Tjarda E; Teunissen, Michelle; van der Veen, Albert J; Emanuel, Kaj S; Tryfonidou, Marianna A; Meij, Bjorn P
(2021) Veterinary Surgery, volume 50, issue 5, pp. 1087 - 1097
(Article)
Abstract
Objective: To evaluate the change in ex vivo biomechanical properties of the canine cervical spine, due to an intervertebral cage, both as a stand-alone device and in combination with plates. Study Design: Experimental ex vivo study. Animals: Cervical spinal segments (C5-C7) from eight canine cadavers. Methods: The range of motion
... read more
(ROM) and elastic zone stiffness (EZS) of the spines were determined with a four-point bending device in flexion/extension, lateral bending, and axial rotation for four conditions: native, discectomy, cage (at C6-C7), and cage with plates (at C6-C7). The disc height index (DHI) for each condition was determined using radiography. Results: Discectomy resulted in overall increased ROM (p <.01) and EZS (p <.05) and decreased DHI (p <.005) when compared to the native condition. Placement of the cage increased DHI (p <.001) and restored total ROM during flexion/extension, lateral bending and axial rotation, and EZS during flexion/extension to the level of the native spine. Application of the plates further reduced the total ROM during flexion/extension (p <.001) and lateral bending (p <.001), but restored ROM in extension and EZS during lateral bending. No implant failure, subsidence, or significant cage migration occurred during loading. Conclusion: An anchorless intervertebral cage used as a stand-alone device was able to restore the disc height and spinal stability to the level of the native cervical spine, whereas the addition of plates further reduced the spinal unit mobility. Clinical Significance: This study implies that the intervertebral cage may be used as a stand-alone device in the spinal unit fixation in the canine cervical spine.
show less
Download/Full Text
Keywords: Animals, Biomechanical Phenomena, Bone Plates/veterinary, Bone Screws/veterinary, Cadaver, Cervical Vertebrae/surgery, Diskectomy/veterinary, Dog Diseases/surgery, Dogs, Internal Fixators/veterinary, Musculoskeletal Diseases/surgery, Orthopedic Procedures/instrumentation, Radiography, Range of Motion, Articular, Spinal Cord Diseases/surgery, Spinal Fusion/methods, Titanium, General Veterinary
ISSN: 0161-3499
Publisher: Wiley-Blackwell
Note: Funding Information: All authors contributed to the study design, collection and interpretation of data, and writing of the manuscript. The titanium intervertebral cages were donated by DePuy Synthes. The UniLOCK plates and screws were bought at commercial prices from DePuy Synthes in the Netherlands. MAT and BPM are being supported by the Dutch Arthritis Society (LLP22). Publisher Copyright: © 2021 The Authors. Veterinary Surgery published by Wiley Periodicals LLC on behalf of American College of Veterinary Surgeons.
(Peer reviewed)