Validation of a fast, robust, inexpensive, two-tiered neonatal screening test algorithm on dried blood spots for spinal muscular atrophy
Strunk, Annuska; Abbes, Andre; Stuitje, Antoine R.; Hettinga, Chris; Sepers, Eline M.; Snetselaar, Reinier; Schouten, Jan; Asselman, Fay Lynn; Cuppen, Inge; Lemmink, Henny; Ludo van der Pol, W.; Engel, Henk
(2019) International Journal of Neonatal Screening, volume 5, issue 2
(Article)
Abstract
Spinal muscular atrophy (SMA) is one of the leading genetic causes of infant mortality with an incidence of 1:10,000. The recently-introduced antisense oligonucleotide treatment improves the outcome of this disease, in particular when applied at an early stage of progression. The genetic cause of SMA is, in >95% of cases,
... read more
a homozygous deletion of the survival motor neuron 1 (SMN1) gene, which makes the low-cost detection of SMA cases as part of newborn screening programs feasible. We developed and validated a new SALSA MC002 melting curve assay that detects the absence of the SMN1 exon 7 DNA sequence without detecting asymptomatic carriers and reliably discriminates SMN1 from its genetic homolog SMN2 using crude extracts from newborn screening cards. Melting curve analysis shows peaks specific for both the SMN1 gene and the disease modifying SMN2 homolog. The detection of the SMN2 homolog, of which the only clinically relevant difference from the SMN1 gene is a single nucleotide in exon 7, was only used to confirm a correct reaction in samples that lacked the SMN1 gene, and not for SMN2 quantification. We retrieved 47 DBS samples from children with genetically-confirmed SMA, after informed consent from parents, and 375 controls from the national archive of the Dutch National Institute for Public Health and the Environment (RIVM). The assay correctly identified all anonymized and randomized SMA and control samples (i.e., sensitivity and specificity of 100%), without the detection of carriers, on the three most commonly-used PCR platforms with melting curve analysis. This test’s concordance with the second-tier ‘golden standard’ P021 SMA MLPA test was 100%. Using the new P021–B1 version, crude extracts from DBS cards could also be used to determine the SMN2 copy number of SMA patients with a high level of accuracy. The MC002 test showed the feasibility and accuracy of SMA screening in a neonatal screening program.
show less
Download/Full Text
Keywords: Copy numbers SMN1, Melting curve, Newborn screening, SMA, SMN2 gene, Two-tiered testing, Pediatrics, Perinatology, and Child Health, Immunology and Microbiology (miscellaneous), Obstetrics and Gynaecology
ISSN: 2409-515X
Publisher: Multidisciplinary Digital Publishing Institute
Note: Funding Information: This research received no external funding. The authors wish to thank the National Research Committee for Neonatal Screening-RIVM for approval of this clinical validation study and the Reference Laboratory for Neonatal Screening-RIVM and the Dept. for Vaccine Supply and Prevention Programmes-RIVM (RIVM, Bilthoven, The Netherlands) for providing the Dried Blood Spots of the SMA patients and control samples. All MC002 and P021 MLPA kits were provided in kind by MRC-Holland. Publisher Copyright: © 2019 by the authors.