Abstract
A device for recording the EEG during open heart surgery is described. It differs from most standard equipment in two ways. First, the input circuit is completely floating from earth and will withstand 500 V DC. Second, radiofrequency (RF)_shielding and filtering permits continuous recording of the EEG during electro-surgery in